The Pfeiffer ASM 192 T high performance console leak detector provides ultra sensitive detection, reliability and easy operation. The ASM 192 T is a console leak detector for standing operators that can be used in vacuum or sniffing test modes. With a minimum detectable leak rate for helium of 5 · 10-8 in sniffing mode and 5 · 10-13 in vacuum mode.
The ASM 192 T has a powerful oil sealed rotary vane pump with 20 m3/h backing capacity or with a second backing pump 40 m3/h backing capacity. The ASM 192 T is a high performance, console detector with a DN 40 ISO-KF flange. Accessories include: wired or wireless remote control, helium spray gun, sniffer probe, and test chambers.
Pfeiffer ASM 192 T Console Leak Detector Features
- Powerful rotary vane pumps
- 20 m3/h backing capacity
- Can take 2 backing pumps to provide up to 40 m3/h
- backing capacity
- Fast time to test
- Console unit for standing operators
- Minimum detectable leak rate 1 · 10-8 in sniffing mode
- Minimum detectable leak rate 5 · 10-13 in vacuum mode
- Graphic color touch screen user interface
- Easy operation
- Remote Control
- For high production leak testing, test chambers can be mounted on inlet flange
- Ultra sensitive detection limits on large volume components
Pfeiffer ASM 192 T Console Helium Leak Detector Applications
- Coating
- FPD – Flat Panel Display
- Photovoltaics
- Industry
- Isolation vacuums
- Lamp and tube manufacturing
- Heat treatment
- vacuum drying
- vacuum furnaces
- R & D
- Nuclear research
- Fusion technology
- Plasma research
- Particle accelerators
- Space simulation
- Cryogenic research
- Elementary particle physics
- Nanotechnology
Pfeiffer ASM 192 T High Performance Leak Detector Specifications
Pfeiffer Vacuum | Units | ASM 192 T | ASM 192 T2D+ |
---|---|---|---|
Flange | in | DN 40 ISO-KF | DN 50 ISO-KF |
Test Method | Vacuum & Sniffing | Vacuum & Sniffing | |
Detectable gases | 4He | 4He | |
Min detectable leak – vacuum | mbar l/s | ≤ 5 · 10-13 | ≤ 5 · 10-13 |
Min detectable leak – sniffing | mbar l/s | ≤ 1 · 10-8 | ≤ 1 · 10-8 |
Pumping speed for He | l/s | 4.4 | 20 |
Maximum inlet test pressure | hPa | 6 | 30 |
Primary pump type | type | Rotary vane pump | Friction-less ACP pump |
Backing Pump Capacity | m3/h | 20 | 25 |
I/O Interface | RS232, Digital, Analog | ||
Start-up time (20° C) | min | <3 | <3.5 |
Operating Temp | celsius | 10° to 40° | 10° to 40° |
Dimensions | mm | 594 x 692 x 933 | 594 x 692 x 933 |
Weight | kg | 125 | 157 |