VIC MD-490S Helium Leak Detector

VIC MD 490S Portable Dry Helium Leak DetectorThe Vic MD-490S Portable Dry Helium Leak Detector detects leaks for a broad range of applications.  It has helium background suppression, .0.5 second response time, and a maintenance-free dry pump.  This is a portable dry helium leak detector with the most sensitive sniffer available.  A touch-screen graphic user interface (GUI) which allows for greater system control and monitoring during system start up and operation.   The Vic MD-490S Series displays included system status, accept/reject status, system health, setup, calibration and other diagnostics.  Auto-tuning, auto-calibration and auto-zeroing features.   It’s lowest detectable leak rate for helium is 5 x 10-8 mbar l/s.

The Vic MD-490S Portable Dry has an entirely oil-free built in diaphragm fore-pump and 70 liter/second air cooled molecular drag pump.

Vic MD-490S Portable Dry detectors are available to buy or rent.

VIC MD-490S Leak Detector Features & Specifications

  • Portable
  • Maintenance-free, dry pump
  • Fully automatic
  • Helium background suppression
  • Response time is under 0.5 seconds
  • Ready to test in under 3 minutes
  • Minimum detectable helium leak 5.10-1=8 atm.cc/s
  • Fully interactive system diagnostics
  • Front panel key lockout
  • Standard probe and built-in calibrated gas leak
  • Minimum detectable helium leak in sniffing mode 1.10-8 atm.cc/s

VIC MD-490S Leak Detector Industries & Applications:

  • Air Conditioning and Refrigeration Systems
    • coils
    • evaporators
    • condensers
    • compressors
    • hoses
  • Automotive Components
    • radiators
    • air conditioning components
    • fuel tanks
    • fuel injection systems
    • air bags
    • heat exchangers
    • headlights
    • wheels
    • hoses
    • pressure transducers
  • Appliance
    • fire extinguishers
    • air conditioners
    • heat pumps
    • regrigerators
    • freezers
    • gas meters
    • gauges
  • Aerospace components and sub-assemblies
  • Vacuum systems – process gas lines, valves
  • Nuclear – Containment vessels
  • Medical Products and sub-assemblies
  • Food packaging and sub-assemblies
VIC Units  MD 490S
Sensitivity (sniffer)  mbar l/s ≤ 5 · 10-8
Pump 1 type Dry built-in diaphragm fore-pump
Pump 2 type 70-l/s air-cooled molecular drag pump
Calibrated Leak type 150 cc high pressure cylinder
Response Time sec 0.5
Time until ready min ≤3
Dimensions (w x h x d) mm 344 x 395 x 618
Weight kg 31.8

Primary Sidebar

Rental Forms & Warranty

Rental Form

Download our leased equipment terms and conditions.

Warranty

Download our warranty on rebuilt and used leak detectors.

RMA Form

Download our RMA form.

Resources

Unit Conversion Tool

Calculate Flowrate, Leakrate, Pressure & Volume.

Quick RFQ

Quick RFQ

Target Services

  • LINAC Leak Testing
  • Condenser Leak Testing
  • Turbine Leak Testing
  • Pipeline Leak Testing
  • Vacuum Furnace Leak Testing
  • Lyophilizer Leak Testing
  • Leak Detector Rentals
  • Leak Detector Repairs

Helium Leak Detectors

  • Adixen Leak Detectors
  • Alcatel Leak Detectors
  • Agilent Leak Detectors
  • Edwards Leak Detectors
  • Leybold Leak Detectors
  • Pfeiffer Leak Detectors
  • Varian Leak Detectors
  • VEECO/VIC Leak Detectors

© 2026 Siteby DMR