Agilent Varian PD03 Helium Leak Detector

Agilent Varian PR 02 helium leak detectorThe Varian PD03 Helium Leak Detector is a portable unit.  The PD03 is a Helium Mass Spectrometer Leak Detector with fully automated tuning and calibration, a patented Autozero feature,  and a leak indicator display that shows leak rates automatically in a linear or log bar graph, and an audible alarm with dedicated volume control and programmable audio theshold.  A microprocessor-controlled valve operation provides automatic system cycling.

The Varian PD03 uses a dry scroll pump combo with a pumping speed of 3 m3/hr.

 

Varian PD03 Helium Leak Detector Features

  • Fully Automated start up and calibration
  • Dry scroll pump combo
  • Programmable test sequences for improved efficiency
  • High test port pressure allows for detection of large leaks
  • 12 range MDL sensitivity meets the most stringent test requirements
  • High helium pumping speed for fast system response
  • Compact and Lightweight design
  • Hand-held remote performs up to 100 meters indoors
  • Multiple channel remote can be used with up to twenty detectors
  • Interface: RS232 and Analog (standard), Discrete IO (optional)
  • Conforms to CE, UL and CSA standards, assuring global acceptance

Varian PD03 Helium Leak Detector Applications

  • Semiconductor production
    • Process tools
    • Gas panels
    • Gas transfer lines
  • Research and Development
    • Electron microscopes
    • Experimental Chambers
    • Surface analytical systems
    • Space chambers
  • Power Generation
    • Condensers
    • Heat Exchangers
    • Steam circuits
  • Small Parts Manufacturing
    • Automotive
    • Electrical
    • Refrigeration
    • Hermetic packaging
    • Medical & Implantable devices
  • Industrial Process Tools
    • Vacuum Furnaces
    • Electron beam welding
    • Coating systems
  • High energy physics
    • Beam lines
    • Accelerators
    • Synchrotrons

Agilent Varian PD03 Helium Leak Detection Specifications

Agilent Varian VS Series Units PR 02 PD 03 BD/MD 15
Configuration options  – Portable Portable Bench/Mobile
Primary pump type type Rotary vane pump Dry scroll pump Dry scroll pump
Pump model DS-42 Dry pump combo IDP-15
Pumping speed m3/hr 2 3 15
Min detectable leak mbar l/s ≤ 5 · 10-12 ≤ 5 · 10-12 ≤ 5 · 10-12
Maximum inlet pressure mbar 13 13 13
Helium pumping speed l/s 1.8 1.8 1.8
Measurement ranges decades 12 12 12
Time until ready min ≤2 ≤2 ≤2
Response time sec <0.5 <0.5 <0.5
Dimensions (w x h x d) mm 567 x 441 x 396 567 x 441 x 396 567 x 441 x 396

How to Rent a Varian PD03 Helium Leak Detector

Related Data Sheets and Documents

Agilent_VS_Brochure

Agilent_VS Series Helium Mass Spectrometer Leak Detectors Operations Manual

Primary Sidebar

Rental Forms & Warranty

Rental Form

Download our leased equipment terms and conditions.

Warranty

Download our warranty on rebuilt and used leak detectors.

RMA Form

Download our RMA form.

Resources

Unit Conversion Tool

Calculate Flowrate, Leakrate, Pressure & Volume.

Quick RFQ

Quick RFQ

Target Services

  • LINAC Leak Testing
  • Condenser Leak Testing
  • Turbine Leak Testing
  • Pipeline Leak Testing
  • Vacuum Furnace Leak Testing
  • Lyophilizer Leak Testing
  • Leak Detector Rentals
  • Leak Detector Repairs

Helium Leak Detectors

  • Adixen Leak Detectors
  • Alcatel Leak Detectors
  • Agilent Leak Detectors
  • Edwards Leak Detectors
  • Leybold Leak Detectors
  • Pfeiffer Leak Detectors
  • Varian Leak Detectors
  • VEECO/VIC Leak Detectors

© 2026 Siteby DMR